Dampval

Library Items

Occasional pipette performance test

Dampval

In a professional testing approach, measurement errors caused by evaporation must be kept to an absolute minimum making the use of evaporation traps indispensable. With evaporation measurements under standardized conditions, we could establish that the evaporation rate with the XP/XS evaporation trap was 59 times smaller compared to an open beaker. This accessory is recommended by the ISO8655 norm for pipette calibration.

It is very easy to convert your balance to a pipette calibration station by simply mounting the corresponding evaporation trap.

   
Evaporation trap for analytical balance XP/XS/XA  
Evaporation trap for MCP system  
Evaporation trap for microbalance XP26 & XP56  
Evaporation trap for NewClassic MS Semi-Micro balances  
Evaporation trap for WXT & WXS systems  